Calibrating module for local hyperthermia complex «Phoenix-2»

Download article in PDF format

Authors: Han K. I., Semenov V. D., Pahmurin D. O., Kobzev A. V.

Annotation: The objective of calibrating module is to heat needle heater and maintain its temperature on certain level with maximum deviation ±0.1 °C. The purpose is achieved using precision temperature sensor and proportional-integral-derivative controller in automatic control system. According the test results, the static error does not exceed 0.04 °C.

Keywords: temperature stabilization, temperature, automatic control system